Standard Practice for Use of Calorimetric Dosimetry Systems for Electron Beam Dose Measurements and Routine Dosimeter Calibration

$83.00

$83.00

$149.40


Sub Total (1 Item(s))

$ 0.00

Estimated Shipping

$ 0.00

Total (Pre-Tax)

$ 0.00


Document Preview Not Available...

1.1 This practice covers the preparation and use of semi-adiabatic calorimeters for measurement of absorbed dose in graphite, water, or polystyrene when irradiated with electrons. The calorimeters are either transported by a conveyor past a scanned electron beam or are stationary in a broadened beam. It also covers the use of these calorimeters to calibrate dosimeter systems in electron beams intended for radiation processing applications.

1.2 This practice applies to electron beams in the energy range from 4 to 12 MeV.

1.3 The absorbed dose range depends on the absorbing material and the irradiation and measurement conditions. Minimum dose is approximately 100 Gy and maximum dose is approximately 50 kGy.

1.4 The averaged absorbed dose rate range shall generally be greater than 10 Gys1, but depends on the same conditions as above.

1.5 The temperature range for use of these calorimeters depends on the thermal resistance of the materials and on the calibration range of the temperature sensor.

1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.


SDO ASTM: ASTM International
Document Number ISO
Publication Date Jan. 22, 2002
Language en - English
Page Count 9
Revision Level 02
Supercedes
Committee E10.01
Publish Date Document Id Type View
May 15, 2019 E1631-20E01 Revision
May 15, 2019 E1631-20 Revision
March 1, 2013 E1631-13 Revision
May 28, 2003 E1631-03 Revision
Jan. 22, 2002 E1631-02 Revision