Standard Practice for X-ray Diffraction Determination of Phase Content of Plasma-Sprayed Hydroxyapatite Coatings

$73.00

$73.00

$131.40


Sub Total (1 Item(s))

$ 0.00

Estimated Shipping

$ 0.00

Total (Pre-Tax)

$ 0.00


...

1.1 This practice is for the determination, by the Reference Intensity Ratio External Standard Method, of the percent by weight of the crystalline phases, hydroxyapatite (HA), beta-(whitlockite) tricalcium phosphate (B-TCP), and calcium oxide (CaO) in coatings deposited upon metallic substrates by plasma-spraying hydroxyapatite.

1.2 A major component in plasma-sprayed HA coatings other than HA is expected to be amorphous calcium phosphate (ACP). Crystalline components other than HA that may be present include alpha- and beta- (whitlockite) tricalcium phosphates, tetracalcium phosphate (TTCP), calcium oxide, and calcium pyrophosphates. Quantification of the minor crystalline components has proven to be very unreliable due to extreme overlap and confounding of X-ray diffraction peaks. Therefore, this practice addresses the quantification of only HA, B -TCP, and CaO.

1.3 This practice was developed for plasma-sprayed HA coatings with HA contents of at least 50 % of the total coating. It is recognized that the analysis of the crystalline components uses diffraction from regions of the pattern that also includes a small contribution from the amorphous component. However, within the limits of applicability of this practice, the effect of such interference is believed to be negligible.

1.4 The coating analyzed shall be produced and processed under equivalent manufacturing conditions to that on the device of interest.

1.5 This practice requires the use of monochromated copper K radiation and flat samples.

This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.


Calcium phosphate coatings have been shown in animal and clinical studies to be biocompatible and to enhance the early attachment of bone to implant surfaces (see Refs. 1-5)4

It is believed that the form of calcium phosphate ceramic and its purity with respect to secondary crystalline phases and amorphous material have an effect on its physical, mechanical, and biological properties. However, no definitive studies of effects on biological properties have been completed. To achieve reproducible clinical results and to permit the determination of the effects of properties of the coating on biological performance, it is essential that the properties of both clinical and experimental materials be well-characterized and consistent.

This practice provides procedures for determination of the percentage by weight of the crystalline phases identified as hydroxyapatite, β-TCP and CaO in plasma-sprayed hydroxyapatite coatings.

SDO ASTM: ASTM International
Document Number F2024
Publication Date May 10, 2000
Language en - English
Page Count 4
Revision Level 00
Supercedes
Committee F04.13
Publish Date Document Id Type View
June 1, 2010 F2024-10 Revision
May 10, 2000 F2024-00 Revision
Feb. 1, 2021 F2024-10R21 Reaffirmation
Oct. 1, 2016 F2024-10R16 Reaffirmation